P****4 发帖数: 831 | 1 最近做SPME-GC实验搞断了几根fiber,很郁闷。
发现问题是GC inlet的septum很难用SPME syringe穿过,好不容易穿过后,septum的碎
屑沾在syringe头上,当脆弱的fiber出来的就断了。但是如果事先将septum的洞戳大的
话,又很可能会漏气。我在考虑是不是我的septum不适合SPME。我现在用Thermo
Scientific的Septa coated BTO 17 mm. 请问各位做SPME-GC的达人一般都用什么fiber
? 或是给我的问题提点建议?请各位大牛多多指教,谢谢先! |
l*****e 发帖数: 3343 | |
l*****e 发帖数: 3343 | 3 Molded Thermogreen™ LB-2 Septa, with injection hole |
s**q 发帖数: 177 | 4 I never had any problems that you talked above. I guess you didn't buy the
right liner and septumS. Check Spelco website and make sure you use the
right parts. |
P****4 发帖数: 831 | 5 Thanks a lot!
I will order this type of Septa to try again. |