d********i
发帖数: 70
1
买了一个二手的,重组装的,但是问题是manual injector搬到inject后,peek tubing
就pop out,特别是从pump出来的那头.是不是不应该用peek材料的.
大侠指导一下吧
K*********e
发帖数: 183
2
As long as pressure is not very high when you run solvent through column,
PEEK tubing or ferrules should not be the problem.
Try to unplug it and tight it in right way again.
If this happen before you even inject anything again, I think the
reassembly might have problem from your vendor.
If it happen when you inject, then it might have particle blocked somewhere.
Just my opinion. Not "da xia".

【在 d********i 的大作中提到】
: 买了一个二手的,重组装的,但是问题是manual injector搬到inject后,peek tubing
: 就pop out,特别是从pump出来的那头.是不是不应该用peek材料的.
: 大侠指导一下吧

k****s
发帖数: 66
3
你压力调的多少?
o****e
发帖数: 1011
4
PEEK材料没问题。
没line堵的话,就尽量拧紧一些。(别拧断就好)
尤其拧进的时候,注意nut,line和ferrule的配合。

【在 d********i 的大作中提到】
: 买了一个二手的,重组装的,但是问题是manual injector搬到inject后,peek tubing
: 就pop out,特别是从pump出来的那头.是不是不应该用peek材料的.
: 大侠指导一下吧

1 (共1页)
进入Chemistry版参与讨论
相关主题
跪求.review opportunitiesused hplc
请问在萃取的时候,为什么我很多时候都看见有两种solvent一起萃取呢Please help me with this paper. I could not download it. Please help me with this paper. I could not
请教几个NMR问题实验室 solvent recovery system
Paper help!HPLC 的injector的Syringe是由什么来控制的呢
ultrafilter请问怎么买?包子autosampler 老是吸不上样品
求建议:Waters HPLC lost prime请问一个HPLC的问题
HPLC杂质峰请教HPLC
请教nucleoside的phosphorylation,包子酬谢Why big variance on LC/MS
相关话题的讨论汇总
话题: peek话题: hplc话题: inject话题: tubing话题: happen