x*****3 发帖数: 3236 | 1 家需要做点renovation,因为孩子还很小,家里老人都非常concerned,觉得会对孩子的
健康很不好(国内好像很少人住着做renovation的),想问问有没有仪器可以自己用来测
空气质量,尤其看看例如甲醛这种有害气体有没有超标。
谢谢。
★ 发自iPhone App: ChineseWeb 7.5 | b********t 发帖数: 1500 | 2
GC, GC/FID, GC/MS
or spectrometers
http://www.picarro.com/gas_analyzers/formaldehyde
or EPA METHOD 8520
http://www.epa.gov/wastes/hazard/testmethods/sw846/pdfs/8520.pd
or EPA method 0100
http://www.epa.gov/wastes/hazard/testmethods/sw846/pdfs/0100.pd
or Detector Glass Tube and pump
device: Detector Tube
manufacturer: Matheson-Kitagawa
model/type: 8014-171SB
sampling information: follow manufacturer’s instructions
upper measurement limit: 35 ppm
detection limit: approx. 1 ppm
overall uncertainty: unknown
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Matheson-Kitagawa
model/type: 8014-171SA
sampling information: follow manufacturer’s instructions
upper measurement limit: 1500 ppm
detection limit: approx. 20 ppm
overall uncertainty: unknown
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Matheson-Kitagawa
model/type: 8014-171SC
sampling information: follow manufacturer’s instructions
upper measurement limit: 4 ppm
detection limit: approx. 0.1 ppm
overall uncertainty: unknown
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: AUER/MSA
model/type: Formaldehyde-0,1 MSA P/N 497649, AUER P/N 5086-813
sampling information: follow manufacturer’s instructions
upper measurement limit: 55 ppm
detection limit: approx. 0.1 ppm
overall uncertainty: unknown
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Gastec
model/type: 91
sampling information: 0.5 to 2 strokes
upper measurement limit: 100 ppm
detection limit: approx. 0.5 ppm
overall uncertainty: 25% for 2 to 6 ppm, 16.4% for 6 to 20 ppm
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Gastec
model/type: 91M
sampling information: 0.5 to 2 strokes
upper measurement limit: 6400 ppm
detection limit: approx. 4 ppm
overall uncertainty: 16.4% for 8 to 200 ppm, 8.2% for 500 to 2000 ppm
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Gastec
model/type: 91L
sampling information: 1 to 2 strokes
upper measurement limit: 40 ppm
detection limit: approx. 0.05 ppm
overall uncertainty: 16.4% for 0.1 to 0.5 ppm, 8.2% for 0.5 to 5 ppm
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Gastec
model/type: 91LL
sampling information: 5 strokes
upper measurement limit: 1.0 ppm
detection limit: approx. 0.03 ppm
overall uncertainty: 16.4% for 0.05 to 0.3 ppm, 8.2% for 0.5 to 1 ppm
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Dräger
model/type: Formaldehyde 0.2/a, order no. 67 33081
sampling information: 10 or 20 strokes
upper measurement limit: 5 ppm
detection limit: approx. 0.1 ppm
overall uncertainty: 33%
method reference: on-site air secondary (manufacturer)
device: Detector Tube
manufacturer: Dräger
model/type: Formaldehyde 2/a, order no. 81 01751
sampling information: 5 strokes
upper measurement limit: 40 ppm
detection limit: approx. 0.5 ppm
overall uncertainty: 33%
method reference: on-site air secondary (manufacturer)
【在 x*****3 的大作中提到】 : 家需要做点renovation,因为孩子还很小,家里老人都非常concerned,觉得会对孩子的 : 健康很不好(国内好像很少人住着做renovation的),想问问有没有仪器可以自己用来测 : 空气质量,尤其看看例如甲醛这种有害气体有没有超标。 : 谢谢。 : ★ 发自iPhone App: ChineseWeb 7.5
| p*****e 发帖数: 537 | | x*****3 发帖数: 3236 | 4 谢谢楼上回答。自己在网上作了点research,考虑买个air purifier用。不知道本版有
木有用过的前辈来说说? |
|