h******o 发帖数: 976 | 1 A nurse caring for a client with a pneumothorax and who has had a
chest tube inserted notes continuous gentle bubbling in the suction
control chamber. What action is appropriate?
1. Do nothing, because this is an expected finding.
2. Immediately clamp the chest tube and notify the physician.
3. Check for an air leak because the bubbling should be
intermittent.
4. Increase the suction pressure so that the bubbling becomes
vigorous.
选什么?我选3,答案却是1,解释起来跟之前理论上说的是相反的... |
j**********n 发帖数: 63 | |
c********i 发帖数: 586 | 3 1 is correct.
suction chamber--should be continously bubbling
water seal chamber--should bubble intermittently "tidaling"
if continously bubbling, air leak
drainage chamber--should not bubble at all |
c********i 发帖数: 586 | 4 Saunders comprehensive review 4th
page 260 |
h******o 发帖数: 976 | 5
thanks!
【在 c********i 的大作中提到】 : 1 is correct. : suction chamber--should be continously bubbling : water seal chamber--should bubble intermittently "tidaling" : if continously bubbling, air leak : drainage chamber--should not bubble at all
|